OmniCure LS200 UV LED Radiometry and Calibration System

OmniCure LS200 UV LED Radiometry and Calibration System

The OmniCure LS200 UV LED radiometry and calibration system ensures precise measurement and calibration of UV LED power or irradiance directly at the cure site, enabling manufacturers to control light output for a reliable and repeatable curing process. The LS200 features seamless integration with the LX500 UV LED pot Curing System for calibration, and offers optional accessories, wide measurement range (1mW to 2000mW or 50mW/cm2 – 40 W/cm2), exceptional precision, and sensor optimization for monochromatic LED sources.

Accurate Measurement and Calibration

Crafted with precision in mind, the LS200 boasts a remarkable accuracy of +/- 10% and a resolution of 1 mW/cm2 for irradiance and 1mW for power. With a fine aperture that is precision calibrated, accurate and consistent measurements can be obtained over a broad dynamic range. The LS200 sensor is factory calibrated to standards traceable to the NRC, ensuring unparalleled accuracy and reliability.

Versatile Measurement Range

Tailor measurements to your specific needs by choosing between irradiance measurement (W/cm2) with the LS200, or optical power measurement (W) with the LS200P. Covering a broad range of power (1 mW to 2000 mW) and irradiance (50 mW/cm2 – 40 W/cm2), and featuring selectable wavelengths programmable from 320-750 nm, the LS200 seamlessly accommodates various applications.  

Compact Design

The LS200 sensor has been designed to take measurements directly at the curing site in even the most confined areas. The sensor can be seamlessly linked to either the LM2011 light meter or the LX500 controller for real-time viewing of optical measurements.

Automation Capability

The LM2011 light meter is equipped with a USB port, enabling communication between the LS200 light sensor and a PC through a user-friendly GUI or an alternative command-line interface to simplify integration into automated workflows.

Optimized for UV LED Curing Systems

Calibrated with a near monochromatic source, the LS200 eliminates the potential for inaccuracies in measurement that can otherwise occur due to the narrow spectral distribution of UV LED sources.

Paired with the OmniCure Beam Positioning Kit, the LS200 seamlessly integrates with the LX500 controller, delivering unmatched control and repeatability in output measurements of OmniCure UV LED Heads.

Additional Accessories

Explore the wide array of OmniCure calibration kits, featuring essential components such as Light Sensors, the LM2011 Light Meter Display for spot checks, and the Beam Positioning Kit for accurate LED head alignment. Combining an accurate radiometer with a precise alignment fixture allows users to efficiently measure the peak irradiance accurately and consistently on every measurement in demanding manufacturing environments. Streamlining setup, these kits provide all the necessary tools in one package for accurate measurement and calibration.

   

Spectral Sensitivity

Selectable wavelengths programmable from 320-750 nm

Measurement Range

Irradiance: 50 mW/cm2 to 40 W/cm2
Power: 1 mW to 2000 mW

Accuracy

+/- 10%

Resolution

Irradiance: 1 mW/cm2
Power: 1 mW

Calibration Cycle

One Year

Battery Life

Two Years, Typical (intermittent use)

Operating Temperature Range

Ambient to 40° C

Certification

Complies with IEC, Canadian and US Standards. RoHS compliant.

Functions

Peak Hold, Power On/Off, Wavelength selection, Power/Irradiance

Calibration

NRC Traceable

LS200 Sensor Dimension (LxWxH)

75 mm x 25 mm x 11.95 mm

LM2011 Light Meter Dimension (LxWxH)

165 mm x 100 mm x 44 mm

   

Spectral Sensitivity

Selectable wavelengths programmable from 320-750 nm

Measurement Range

Irradiance: 50 mW/cm2 to 40 W/cm2
Power: 1 mW to 2000 mW

Accuracy

+/- 10%

Resolution

Irradiance: 1 mW/cm2
Power: 1 mW

Calibration Cycle

One Year

Battery Life

Two Years, Typical (intermittent use)

Operating Temperature Range

Ambient to 40° C

Certification

Complies with IEC, Canadian and US Standards. RoHS compliant.

Functions

Peak Hold, Power On/Off, Wavelength selection, Power/Irradiance

Calibration

NRC Traceable

LS200 Sensor Dimension (LxWxH)

75 mm x 25 mm x 11.95 mm

LM2011 Light Meter Dimension (LxWxH)

165 mm x 100 mm x 44 mm

Download asset:
To download this item, please complete the form below.
NOTE: If you have disabled our marketing cookies, you will not be able to access our videos or downloads. To update your cookie preferences, please click here to review and enable cookies.
CAPTCHA
This question is for testing whether or not you are a human visitor and to prevent automated spam submissions.
Close